Triple-H14 Bag-In/Bag-Out chemotherapy safety cabinets and SUS316L negative/positive-pressure isolators with integrated 6-log flash VHP bio-decontamination.
Guangzhou Hengtai (HENGTAI CLEAN) manufactures two specialized primary engineering controls (C-PEC) for oncology hospitals, 503A/503B compounding pharmacies, ADC/HPAPI biopharma, and cell & gene therapy (CGT): (1) HT-CSC Series DIN 12980 / EN 12469 Triple-HEPA Cytotoxic Safety Cabinets featuring a zero-exposure under-worksurface Bag-In/Bag-Out (BIBO) H14 filter bank, and (2) HT-CACI / HT-CAI Series SUS316L Containment & Aseptic Isolators achieving ISO 10648-2 Class 2 leak tightness (< 0.25% vol/h), 0.45 m/s ±20% Grade A laminar airflow, automated 35% H2O2 VHP sterilization, and 21 CFR Part 11 audit-trail PLC control.

DIN 12980 Inflow Protection: v_inflow ≥ 0.53 m/s across the 200 mm sash aperture + v_downflow = 0.33 m/s unidirectional H14 Grade A stream (100% hard-ducted or thimble-exhausted for volatile HD APIs). ISO 10648-2 Class 2 Pressure Decay: Hourly leak rate Tf = (ΔP / P_atm) × (60 / t_min) ≤ 0.0025 h⁻¹ (0.25% volume/hour) verified at ±250 Pa test pressure prior to shipment.
| Model | Equipment Type & Standard | Internal Work Zone (W×D×H) | Filtration & Pressure Regime | Chamber Material & Decontamination | Engineering BOQ Basis |
|---|---|---|---|---|---|
| HT-CSC-1200 | Cytotoxic Safety Cabinet (DIN 12980 / EN 12469) | 1200 × 620 × 650 mm (2 Operators: 1500/1800 opt.) | Triple H14 HEPA (Downflow + Under-Tray BIBO + Exhaust) · 0.53 m/s Inflow | Seamless SUS316L Coved Workzone · UV-C + Formalin/VHP Port | Custom BOM Quote (24h Itemized BOQ) |
| HT-CSC-1800 | Dual-Operator Oncology Cytotoxic Cabinet | 1800 × 620 × 650 mm | Triple EN 1822 H14 Gel-Seal · Dual EC Blowers · ≤ 56 dB(A) | SUS316L Ra ≤ 0.4 µm · Hard-Duct Canopy Thimble Included | Custom BOM Quote (24h Itemized BOQ) |
| HT-CACI-2G | 2-Glove USP <800> Negative-Pressure Compounding Isolator | 1100 × 700 × 750 mm + 500 mm Pass Chamber | Dual H14 Push-Push BIBO · -35 to -50 Pa Work / -18 Pa Pass | 3.0 mm Orbital-Welded SUS316L · Optional VHP Port | Custom BOM Quote (24h Itemized BOQ) |
| HT-CACI-4G-VHP | 4-Glove OEB 5 HPAPI / ADC VHP Containment Isolator | 1850 × 750 × 800 mm + Dual RTP / VHP Airlocks | Triple H14/U15 BIBO · -50 to -80 Pa · < 1 µg/m³ OEB 5 | SUS316L Ra ≤ 0.38 µm + Built-in 6-Log Flash VHP Generator | Custom BOM Quote (24h Itemized BOQ) |
| HT-CAI-4G-VHP | EU GMP Annex 1 Positive-Pressure Sterile Testing / CGT Isolator | 1850 × 750 × 800 mm + Sterility Pump Integration | Unidirectional Grade A H14 + +50 Pa Positive Cascade | SUS316L + Catalytic < 1 ppm H2O2 Aeration in 35 Min | Custom BOM Quote (24h Itemized BOQ) |
In ordinary Class II biosafety cabinets, contaminated chemotherapy aerosols travel through internal plenums before reaching top filters. Our DIN 12980 HT-CSC places a first-stage H14 BIBO filter bank immediately beneath the perforated SUS316L work tray so all downstream fans and ducts remain 100% uncontaminated.
HT-CACI-4G-VHP and HT-CAI-4G-VHP units incorporate dry flash H2O2 vaporization, Dräger/Vaisala high/low ppm H2O2 sensors, wireless glove-port pressure decay leak testers, and Siemens WinCC Audit Trail logging.
Combine our HT-CSC or HT-CACI units with our pre-engineered 45–65 m² USP <797>/<800> modular hardwall compounding room (-5 to -12 Pa HD buffer room, 30 ACH dedicated exhaust) in a single 40'HQ container.
91440101MA59M8NW4E · Building EPC Grade-II D244805460 · Decoration/Steel/Environmental Grade-II D344227631 · Safety License GD-JZ [2022]010947.GR202444009466 · ISO 9001:2015 90924Q00571R0M · CE & EAC · 63 National Patents & 19 Software Copyrights.
A DIN 12980 cytotoxic cabinet adds a third H14 HEPA filter stage directly underneath the work surface equipped with safe-change Bag-In/Bag-Out (BIBO) housing. This captures hazardous antineoplastic drug particles immediately below the tray so service technicians and internal blowers are never exposed to cytotoxic dust.
Yes. Closed Grade A isolators (HT-CAI / HT-CACI) with validated automated VHP bio-decontamination and RTP/VHP transfer airlocks can operate inside a Grade D (ISO 8) background for aseptic sterility testing or compounding, cutting HVAC energy consumption by over 50% compared to conventional Grade B suites.
We supply wireless automatic Glove Integrity Testers (GIT) with RFID port identification that pressurize each CSM/Hypalon sleeve to +500–800 Pa and detect micropores ≥ 100 µm within a 12-minute pressure-decay cycle prior to batch start.
Every cabinet and isolator undergoes Factory Acceptance Testing (FAT) including EN 1822 PAO HEPA scan, smoke airflow visualization video, ISO 10648-2 pressure decay test, and IQ/OQ English/Russian/Spanish validation binders.
Looking for factory-direct pricing without middleman markups? Send our cleanroom HVAC and stainless equipment engineers your layout, target ISO 14644-1 class, or equipment list.
OEB4/OEB5 operator and product containment for oncology compounding, potent API handling, and sterile filling lines using safe-change push-push HEPA filters and glove-port leak testing.
| Cleanroom Class / Subsystem | Target Standard / Capacity | HENGTAI Factory Specification | Primary Application |
|---|---|---|---|
| Cytotoxic Safety Cabinet | EN 12469 / NSF 49 100% Exhaust | Under-Worktop Safe-Change Bag-Out H14 | Oncology & Hazardous IV |
| Aseptic / Containment Isolator | +50 Pa Aseptic or -75 Pa Potent | Automatic Glove Integrity Tester + VHP | OEB4 / OEB5 (<1 ug/m3 OEL) |
A (HENGTAI Chief Engineer): Zero-exposure Bag-In-Bag-Out (BIBO) or under-plenum safe-change housings seal spent filters inside PVC containment bags before removal.